The limitations first, the price later

You assess whether a method supports the weight put on it. So the limits are at the top of this page and not behind a disclosure triangle.

What a single camera cannot do

Rotation about the long axis of the limb is not recoverable from one view. What is reported as hip adduction in the frontal plane is a projection, and an internally rotated hip changes that projection without changing the adduction. If rotation is the question, the answer is a multi-camera setup, and that is a different installation from a tripod beside a treadmill.

Foot and ankle are the weakest part. The model’s heel and toe points sit on the shoe, not on bone, and a thick midsole shifts the measured foot strike angle systematically. Rearfoot / midfoot / forefoot classification is therefore screening grade and wants calibrating against the footwear in front of you.

Absolute ground contact time from video underestimates by roughly five to ten per cent, because first contact and final toe-off fall between frames. The left-right difference and the change over time are unaffected: that error acts the same way on both sides.

How an event is established

Initial contact underpins nearly every temporal measure in the report. It is not determined by one method but by three, and only counts when at least two agree.

  1. 1

    Foot height

    The minimum of heel and toe against an estimated floor line. Robust, but blunt: near the minimum the foot barely moves, so the exact frame is hard to pick.

  2. 2

    Vertical acceleration

    The peak in the second derivative at landing. This is the sharpest of the three and supplies the timing; the other two only decide whether an event exists.

  3. 3

    Foot relative to pelvis

    The Zeni method: the extreme of the horizontal foot-to-pelvis distance. Works even when the floor is out of frame.

What has and has not been validated

Against a simulated runner with known ground truth: initial contact within zero frames, toe-off within about two, cadence within one step per minute, vertical oscillation within one per cent, and zero asymmetry on a symmetric runner. That tests the arithmetic end to end, not the world.

What does not exist is a study on real footage against force plates or an optoelectronic system. It ought to, and the protocol is written: concurrent measurement against a reference to establish systematic bias per measure; within-session and between-rater repeatability; and from those, the smallest change that reliably clears the noise. Until that exists, this page states what there is and nothing beyond it.

Why this is not a medical device

The software measures and assesses the quality of that measurement. It does not link values to a condition, predict anything, or propose treatment. That places it outside the definition of medical device software, and Rule 11 does not arise.

The reasoning is written out against the five decision steps of MDCG 2019-11 and is a public document. It also states what would change with an interpretation module included: Class IIa, with the route that goes with it. That module was built and deliberately removed again.

Frequently asked

Which reference values are used?
Ranges from the literature, with the caveat that most depend heavily on speed and stature. They sit beside the measurement as orientation, not as cut-offs, and nothing is labelled abnormal.
Can I supply my own norms?
Yes, in the local installation through a norms file. For a service working with a defined population that is more useful than ranges drawn from recreational runners.
Can I get at the raw data?
The measurement exports as JSON with the settings, the scaling used, the cycle count and the spread per measure. Without patient details, so a measurement can be shared or recomputed without a name travelling with it.
Which pose estimator is underneath?
RTMPose with the Halpe26 point set, through onnxruntime. Apache-2.0, so usable without licence restriction, and it runs in the browser. On top of the raw detection sits a correction that holds bone lengths constant across the clip; that removes fifteen to twenty-two per cent of the angular error around foot and ankle.

The evidence page has the rest

How it measures, what has been tested, what has not been tested yet, and the sources underneath.

Read the evidence

A measuring instrument, not a diagnosis. Interpretation rests with the clinician.